Toggle Nav

ISO 9001:2015

90+ U.S. Patents

Since 1954

Tailored For You

Systems / LNG & CNG / LNG Sampling

LNG Sampling Systems

Welker's LNG sampling systems are purpose-built around one fundamental imperative: the total and complete vaporization of the LNG sample — engineered, built, and tested at Welker to ISO 8943: 2025 and API 8.6 standards.

ISO 8943 Compliant API 8.6 / API 14.1 ISO 9001:2015 Certified Cryogenic to -265°F
Welker LNG Sampling System — vaporizer, accumulator, and sample cylinders Welker LNG VAPORIZER System & INTERMITTENT SYSTEM
Total Vaporization. Every Time.

The Most Critical Step in LNG Custody Transfer

LNG is natural gas cooled to approximately -265°F — around 600 times denser than in its gaseous state. Sampling it accurately requires one non-negotiable step: total and complete vaporization before the sample reaches any cylinder or analyzer. Partial vaporization means heavier components separate out, the sample becomes unrepresentative, and custody transfer billing is wrong.

Patented Cryogenic Check Valve Probe Welker's patented probe tip check valve begins vaporization the instant LNG enters — preventing heavier components from returning to the pipeline and eliminating line pack.
Heated Vaporizer — 2,000W Complete Phase Change An electrically heated vaporizer (2,000W / 113 BTU/min) with an impingement block ensures total, controlled vaporization before the sample reaches the accumulator, cylinders, or GC.
Three ISO 8943 System Configurations Welker produces systems aligned with all three ISO 8943 figures — covering every LNG custody transfer scenario from continuous GC analysis to intermittent cylinder collection.
275 psig MAOP · -265°F Rated · NEC Class I Div. 1 316/316L stainless steel construction throughout. Rated to 275 psig MAOP, minimum -265°F inlet temperature. NEC Class I, Division 1, Groups C & D area classification.
Welker Made™ Every LNG sampling system — probe, vaporizer, accumulator, enclosure, cylinders — engineered, manufactured, and tested under one roof by the team that helped write the standard.
How It Works

From Cryogenic Pipeline to Sample in Six Steps

01

Cryogenic Probe Extraction

Welker's patented cryogenic probe draws sub-cooled LNG from the flowing pipeline. The probe tip check valve begins vaporization at the point of entry — preventing compositional separation and line pack before the sample reaches the cabinet.

02

Heat-Traced Insulated Line

Customer-supplied heat-traced, insulated tubing maintains the sample in its sub-cooled state from the probe tip to the cabinet inlet — preserving sample integrity across the critical transfer distance to the vaporizer.

03

Complete Vaporization

The 2,000W heated vaporizer converts the LNG to gas completely and controllably. An impingement block ensures any remaining liquid droplets are fully vaporized before the gas stream continues — no partial vaporization, no fractionation.

04

Accumulator Conditioning

The fully vaporized gas is conditioned in a heated accumulator, which buffers the sample, ensures homogeneity, and provides regulated, stable gas flow to the downstream cylinders or online GC.

05

Cylinder Collection or Online GC

Conditioned gas is delivered to CP-2 Constant Pressure Cylinders or SC Single Cavity Cylinders for lab transport, or directed continuously to an on-site online gas chromatograph for real-time custody transfer measurement .

06

Analysis → Custody Transfer Billing

GC analysis determines BTU content and full LNG composition. Combined with density and quantity transferred, this produces the commercial value of the cargo for GIIGNL-compliant custody transfer billing and documentation.

System Configurator

Find Your LNG Sampling System

Answer two questions about your sampling method and sample destination and we'll point you to the right ISO 8943 configuration.

1. What sampling method does your application require?

2. Where does your vaporized sample need to go?

Recommended System
 
 
Standards & Certifications ISO 8943 — LNG Sampling API 8.6 — LNG Sampling Standard API 14.1 / GPA 2166 NEC Class I, Div. 1, Groups C & D ISO 9001:2015 Certified Manufacturer
System Configurations

Three ISO 8943 LNG Sampling Systems

Every Welker LNG system includes the patented cryogenic probe, heated vaporizer, and accumulator. The configuration differs in how the vaporized sample is collected and delivered.

Welker LNG Vaporizing Sampling System
01 ISO 8943 Fig. 1 · Continuous

Vaporizing System

Continuous sampling → Online GC

The core Welker LNG system. A continuous stream of LNG is drawn through the patented cryogenic probe, completely vaporized in the heated chamber, conditioned in the accumulator, and delivered in real time to an online gas chromatograph.

Key Features

  • Patented probe tip check valve — vaporization begins at probe entry
  • 2,000W heated vaporizer — complete, controlled phase change
  • Heated accumulator — homogeneous, stable gas delivery
  • Continuous gas output to online GC for real-time analysis
  • 316/316L stainless steel

Key Specifications

MAOP275 psig @ -20°F to 120°F Min. Inlet Temp.-265°F (-165°C) Vaporizer Power2,000W (113 BTU/min) Area ClassNEC Class I, Div. 1, C & D

Best Application Fit

LNG terminals Continuous custody transfer Online GC integration Import / export measurement
Welker LNG Vaporizing plus Spot Sampling System
0 SPOT SAMPLING SYSTEM

Vaporizing + Spot Sampling System

Continuous GC analysis + simultaneous cylinder collection

A configuration delivering vaporized LNG to SC Single Cavity Cylinders for laboratory verification. The accumulator splits conditioned gas to both outputs — providing the gold standard for high-value custody transfer where independent dual verification is required.

Key Features

  • Dual output: simultaneous GC + cylinder collection from one vaporizer
  • Online measurement and laboratory backup in a single system
  • Ideal where dual verification is contractually required
  • Same patented probe, vaporizer, and accumulator as System 1
  • SC Single Cavity Cylinders collected as reference samples for dispute resolution
  • Compatible with GIIGNL custody transfer documentation

Key Specifications

MAOP275 psig @ -20°F to 120°F Min. Inlet Temp.-265°F (-165°C) OutputsOnline GC + CP-2 Cylinders Area ClassNEC Class I, Div. 1, C & D

Best Application Fit

High-value LNG cargo Dual verification required FSRU / import terminals Dispute resolution backup
Welker LNG Intermittent CP/FP Sampling System
03 ISO 8943 Fig. 3 · Intermittent CP

Vaporizing + Intermittent Sampling System

Timed spot grabs →CP-2 cylinders → Laboratory GC

The intermittent system collects discrete spot grabs of vaporized LNG into Welker CP-2 Constant Pressure Cylinders at timed intervals throughout the transfer. Sealed cylinders are transported to a laboratory for GC analysis. Welker's CP-2 design is the origin of ISO 8943 Figure 3 — Welker invented the method that became the global standard.

Key Features

  • Welker CP cylinder — the origin of ISO 8943 Figure 3
  • Floating piston maintains sample pressure during transport to lab
  • No on-site GC required — lab-based analysis from sealed cylinder
  • Multiple cylinders collected sequentially for composite representation
  • Pre-cooled cylinder chilling cycle ensures cryogenic sample integrity

Key Specifications

MAOP275 psig @ -20°F to 120°F Min. Inlet Temp.-265°F (-165°C) CollectionCP-2 Constant Pressure Cylinders → Lab GC StandardISO 8943 Fig. 3 (Welker origin)

Best Application Fit

No on-site GC available Lab-based custody transfer Smaller LNG terminals Portable / temporary transfer Requires lab GC access
Side-by-Side Comparison

System Comparison

Use this table to match system configuration against your GC availability, sampling requirements, and custody transfer documentation needs.

Specification System 1 — Vaporizing System 2 — Vaporizing + Spot System 3 — Vaporizing + Intermittent
Sampling MethodContinuousContinuous + SpotIntermittent Spot Grabs
Online GC Output Direct Connection Direct Connection Lab GC only
Cylinder Collection Single Cavity Cylinders✔  Constant Pressure Cylinders
Real-Time Custody Data Lab results after transfer
Patented Cryogenic Probe
Heated Vaporizer (2,000W)
Heated Accumulator
On-Site GC Required For GC output For GC output
MAOP275 psig275 psig275 psig
Construction316/316L SS316/316L SS316/316L SS

Ready to Configure Your LNG Sampling System?

Our LNG sampling engineers will help you select the right configuration for your terminal, GC availability, custody transfer requirements, and sampling standards.

Contact Us